To remove and replace the cover

Chapter 6: Replacing Assemblies

To remove and replace the cover

To remove and replace the cover

Use these steps to remove and replace the cover. When necessary, refer to other removal procedures.

1Disconnect the power cable.

2Disconnect all scope probes and BNC input cables from the front panel.

3Disconnect any other cables, such as mouse, keyboard, printer, or GPIB cables.

4 Remove the two Torx T15 screws securing the side handle.

5 Remove the four Torx T15 screws that secure the rear feet (two in each foot). 6 Remove the four Torx T15 screws that secure the cover to the chassis.

7Set the unit on its side. Carefully slide the cabinet off of the frame by pulling the front panel and the cover away from each other.

8To replace the cover, reverse the above procedure.

Be sure to keep ribbon cables out of the way when replacing the cover, particularly the flex cable and connector for the AutoProbe assembly at the bottom front of the oscilloscope.

CAUTION

PROPERLY TIGHTEN HANDLE AND SCREWS!

 

Tighten the side handle and rear feet screws to 2 Nm (18 in-lbs).

 

 

Figure 6-1

 

Cover Fasteners