10Select Acquisition from the Setup menu.

Chapter 3: Testing Performance

To test time measurement accuracy

8Select horizontal from the setup mean. Set the scale to 5 ns/div, position at –5 ns, and reference to the left.

Figure 3-8

Horizontal Setup for Equivalent Time Procedure

9Adjust the signal generator output voltage to obtain a waveform with a risetime of approximately 700 ps to 1.4 ns.

You can measure the risetime by selecting the Risetime command from the Time submenu of the Measure menu. See figure 3-9.

Figure 3-9

Use the risetime measurement to verify correct waveform shape for the time measurement accuracy check

Waveform for Time Measurement Accuracy Check

10Select Acquisition from the Setup menu.