Chapter 3: Testing Performance

To test voltage measurement accuracy

To test voltage measurement accuracy

This test verifies the voltage measurement accuracy of the instrument. The measurement is made using dual-cursor automatic measurement so that offset errors are not a factor.

A power supply provides a reference voltage to check voltage measurement accuracy. The actual supply voltage is monitored for accuracy using a NIST-traceable voltmeter. A dc blocking capacitor and a BNC short are used to filter any noise generated by the power supply (especially at low voltages) so it does not appear at the oscilloscope input.

Specification

Only the dual-cursor specification is tested.

Single Cursor Measurement: ±(gain accuracy + offset accuracy + resolution/2)

Dual Cursor Measurement: ±(gain accuracy + resolution)

Gain Accuracy: ± 1.00% of full scale at full-resolution channel scale

Resolution: 8 bits, (0.4% of full scale without averaging); or

12 bits, (0.024% of full scale with 32 averages) at ful-resolution scale Offset Accuracy: ±(1.00% of channel offset + 1% of full scale) at full-resolution scale

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Power Supply

7 mV to 30 Vdc, 0.1 mV resolution

Agilent 6114A

Digital Multimeter

Better than 0.1% accuracy

Agilent 34401A

(DVM)

 

 

Cables (2)

BNC

Agilent 10503A

Adapters (2)

BNC (f) to banana (m)

Agilent 1251-2277

Adapters (2)

BNC tee (m)(f)(f)

Agilent 1250-0781

Blocking capacitor

0.18 F

Agilent 10240B

Shorting cap

BNC

Agilent 1250-0774

Procedure

1Connect the equipment.

See figure 3-4.

a Use a BNC-to-banana adapter to connect a BNC cable to the power supply.

b Connect a BNC tee to the other end of the cable and connect the tee to channel 1 of the oscilloscope.

c Connect another BNC cable to the tee at the oscilloscope and connect a BNC tee to the other end of the cable.

d Connect the blocking capacitor to the BNC tee and connect the BNC short to the blocking capacitor.

e Connect a BNC-to-banana adapter to the same BNC tee and connect the adapter to the DVM input.

3–7

Page 44
Image 44
Agilent Technologies 54835A, 45A, 46A manual To test voltage measurement accuracy, Connect the equipment, Specification