Agilent Technologies 45A, 46A manual To check the flat panel display FPD, Calibrating and Adjusting

Models: 54835A 46A 45A

1 207
Download 207 pages 4.15 Kb
Page 72
Image 72
To check the flat panel display (FPD)

Chapter 4: Calibrating and Adjusting

To check the flat panel display (FPD)

To check the flat panel display (FPD)

No equipment is required for this procedure. Specifications for flat-panel displays used in the Infiniium oscilloscope are shown in the following table.

Flat-Panel Display Specifications

Defect Type

Polarizer

Dot Defect

(A dot is defined as 1, 2, or 3 stuck subpixels touching horizontally. Subpixels are horizontal red, green, blue triads, so these may show up as one of 8 colors or black.)

Line Defect

Non-uniformity

Luminance

(The measurement is perpendicular to the screen surface in both axes.)

 

Limit

Scratch

Width 0.05 mm

 

Length 10 mm

Dent

φ ≤ 0.4mm

Bright dot

N 5

Dark dot

N 7

Total dot

N 12

Two adjacent dots

2 pairs

Bright dot

Dark dot

2 pairs

Three or more adjacent dots

Not allowed

Distance between defects

10 mm

Bright dot

Dark dot

10 mm

 

Not allowed

 

Check other specifications

Minimum

160 cd/m2

Typical

200 cd/m2

When to Use this Procedure

This procedure should not be performed as a part of routine maintenance. Perform the procedure only when there appears to be a problem with the display.

1Enable the graphical interface.

2Select Self Test from the Utilities menu.

3Click the Service Extensions box in the Self Test dialog.

4 Select Screen from the Test drop-down list box.

See figure 4-2.

4–7

Page 72
Image 72
Agilent Technologies 45A To check the flat panel display FPD, Calibrating and Adjusting, Flat-Panel Display Specifications