Chapter 6: Replacing Assemblies

To remove and replace an acquisition hybrid

 

To remove and replace an acquisition hybrid

 

Use the following procedure to remove and replace an acquisition hybrid. When necessary, refer

 

to other removal procedures.

 

 

CAUTION

ELECTROSTATIC DISCHARGE!

 

Use grounded wrist straps and mats when servicing the acquisition board. Electrostatic discharge

 

can damage electronic components.

 

You do not have to remove the acquisition board before replacing an acquisition hybrid.

 

 

To understand the sequence of parts, use the accompanying illustration.

To remove the acquisition hybrid

1Disconnect the power cable and remove the cover.

2Use a T-6 Torx driver to remove two screws that secure the heatsink spring, then remove the heatsink.

3Disconnect the semi-rigid cables from the A/D-FISO hybrids, then disconnect from the attenuator ends.

4First, use a T-6 Torx driver to remove the four screws that secure the top plate.

5 Use a 3/16 hex driver to remove two standoffs that secure the top plate. 6 Remove the top plate.

7Grasp the hybrid across the diagonal corners (or sides) using your thumb and index finger. Then lift the hybrid off of the connector assembly.

6–32