Chapter 6: Replacing Assemblies

To remove and replace an attenuator

 

To remove and replace an attenuator

 

Use this procedure to remove and replace an attenuator assembly. When necessary, refer to

 

other removal procedures.

 

 

CAUTION

ELECTROSTATIC DISCHARGE!

 

Use grounded wrist straps and mats when servicing the acquisition board. Electrostatic discharge

 

can damage electronic components.

 

 

Do Not Remove the Attenuators from the Acquisition Board

Attenuators are part of the acquisition board. When replacing the acquisition assembly, do not remove the attenuators.

1Using ESD procedures, remove the acquisition board assembly from the unit.

2Using a Torx T6 driver, remove the heatsinks from the associated A/D-FISO hybrids.

The heatsinks must be removed first to allow you to disconnect the semi-rigid cables from the hybrid end before disconnecting the cables from the attenuator end.

3Carefully disconnect the semi-rigid cables from the A/D-FISO hybrids, then disconnect from the attenuator ends.

4Carefully lift up on the rear of the attenuator boards near the two mating connectors to remove from the acquisition board.

5Replace the attenuator by reversing this procedure. Torque the semi-rigid cables to no more than 5 in-lbs and the heatsink screws A13H3 to 3 in-lbs.

6–28

Page 165
Image 165
Agilent Technologies 45A To remove and replace an attenuator, Do Not Remove the Attenuators from the Acquisition Board