To check the 715 Hz auxiliary output (probe compensation squarewave)

Chapter 4: Calibrating and Adjusting

To check the 715 Hz auxiliary output (probe compensation squarewave)

To check the 715 Hz auxiliary output (probe compensation squarewave)

This test is optional. The 715 Hz auxiliary output is not specified in the oscilloscope performance specifications. The values given are typical. Results are not recorded in the Performance Test Record.

Equipment Required

You can check the 715 Hz auxiliary output using the scope itself.

Procedure

1Press Default Setup.

2Use a BNC cable to connect the front panel Aux Out to the Channel 1 input.

3 Set Coupling to DC and Input to 50for Channel 1.

4 Press Autoscale.

5 Select Vamptd from the Voltage submenu of the Measure menu. 6 Select Frequency from the Time submenu of the Measure menu.

7The signal should be a squarewave at approximately 715 Hz with amplitude of approximately 300 mVpp.