The following tables list the performance specifications and operating characteristics for the Agilent Technologies 54835A/45A/46A oscilloscope. Asterisks (*) denotes warranted specifications, all others are typical. Specifications are valid after a 30 minute warm-up period, and within ± 5 °C from the self-calibration temperature.

Acquisition

Maximum Sample Rate

Real Time

Agilent Models

 

 

54835A — 2-channel mode: 4 GSa/s

 

 

54835A — 4-channel mode: 2 GSa/s

 

 

54845A — 2-channel mode: 8 GSa/s

 

 

54845A — 4-channel mode: 4 GSa/s

 

 

54846A — 2-channel mode: 8 GSa/s

 

 

54846A — 4-channel mode: 4 GSa/s

Maximum Effective Sample Rate

Equivalent Time

500 GSa/s

Memory Depth

2-channel mode: 65,536 points

Agilent 54835A, Agilent 54845A, and

 

4-channel mode: 32,768 points

Agilent 54846A

Memory Depth Modes

Auto

Optimized for best combination of

 

 

update rate and display quality.

 

Manual

Selectable

 

 

2-channel mode: from 16 to 65,536

 

 

points

 

 

4-channel mode: from 16 to 32,768

 

 

points

Sampling Modes

Real Time

Successive single shot acquisitions.

 

Equivalent Time

Random repetitive sampling (higher

 

 

time resolution at fast sweep speeds).

Filters

9-bit Bandwidth Limit filter: BW = (Sample Rate)/20

 

(Sin x)/x Interpolation: On/Off selectable FIR digital filter. Digital signal

 

processing adds points between acquired data points to enhance

 

measurement accuracy and waveform display quality.

 

BW= Sample Rate/4

 

Averaging

Selectable from 2 to 4096.

 