Chapter 6: Replacing Assemblies

To remove and replace the AutoProbe assembly

cFrom the back of the front panel, use a small screw driver or other slender pointed object through the access hole to push the AutoProbe assembly faceplate away from the front panel assembly.

Figure 6-6

 

Pushing Out the AutoProbe Faceplate

 

 

CAUTION

AVOID DAMAGE TO THE RIBBON CABLE AND FACEPLATE!

 

Do not pry around the edge of the assembly. Doing so may damage the ribbon cable or faceplate.

 

 

dFinish removing the AutoProbe assembly by feeding the mylar ribbon cable through the front panel slot.

3To replace the AutoProbe assembly, reverse the above procedure.

6–7

Page 144
Image 144
Agilent Technologies 45A, 46A, 54835A Avoid Damage to the Ribbon Cable and Faceplate, Pushing Out the AutoProbe Faceplate