To connect power

Chapter 2: Preparing for Use

To connect power

To connect power

The instrument Power Factor Correction (PFC) circuitry in the oscilloscope’s power supply operates over a line voltage in the range of 100 to 240 Vac ±10% (the power supply is autoranging to the input voltage and frequency). Line frequency must be in the range 47 to 440 Hz. Power consumption is 390W maximum.

1Position the instrument where it will have sufficient clearance for airflow around the top, back, and sides.

Minimum 0 mm

 

Minimum 38.1 mm

 

 

 

 

 

 

 

 

 

 

 

 

Minimum 15.9 mm (maintained by feet; however, do not place on soft surfaces that will impede airflow)

Airflow requirements

Agilent 54835A/45A/46A – 250

Minimum A Three-wire Power Cable is Provided

15.9mm

both sides

2Connect the power cord to the rear of the oscilloscope, then to a suitable ac voltage source.

 

 

The line cord provided is matched by Agilent Technologies to the country of origin of order.

 

 

Ensure that you have the correct line cord. See “Power Cables and Plug Configurations” in

 

 

chapter 7.

 

 

 

WARNING

 

SHOCK HAZARD!

 

 

BEFORE YOU CONNECT THIS INSTRUMENT TO MAINS POWER OR LIVE MEASURING

 

 

CIRCUITS, you must provide a protective earth ground. Failure to provide a protective earth

 

 

ground could result in a shock hazard if there is a failure in this instrument or equipment

 

 

connected to it.

 

 

The mains plug must be inserted in a socket outlet provided with a protective earth contact.

 

 

Do not use an extension cord (power cable) without a protective conductor (grounding).

 

 

Grounding one conductor of a two-conductor outlet does not provide an instrument ground.

 

 

 

 

 

 

 

 

A Three-wire Power Cable is Provided

 

 

This instrument is provided with a three-wire power cable. When connected to an appropriate ac power

 

 

outlet, this cable grounds the instrument cabinet.

 

 

 