To test offset accuracy

Chapter 3: Testing Performance

To test offset accuracy

To test offset accuracy

This test checks the vertical offset accuracy.

Specification: ±(1.00% of channel offset + 1% of full scale) at full-resolution channel scale

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Power Supply

0.5 V to 2 Vdc, ±1 mV accuracy

Agilent 6114A

Digital Multimeter

Better than 0.1% accuracy

Agilent 34401A

(DVM)

 

 

Cables (2)

BNC

Agilent 10503A

Adapters (2)

BNC (f) to banana (m)

Agilent 1251-2277

Adapters (2)

BNC tee (m)(f)(f)

Agilent 1250-0781

Blocking capacitor

0.18 F

Agilent 10240B

Shorting cap

BNC

Agilent 1250-0774

Procedure

1Connect the equipment.

The cabling is the same as that used for the voltage measurement accuracy test. See figure 3-4. a Use a BNC-to-banana adapter to connect a BNC cable to the power supply.

b Connect a BNC tee to the other end of the cable and connect the tee to channel 1 of the oscilloscope.

c Connect another BNC cable to the tee at the oscilloscope and connect a BNC tee to the other end of the cable.

d Connect the blocking capacitor to the BNC tee and connect the BNC short to the blocking capacitor.

e Connect a BNC-to-banana adapter to the same BNC tee and connect the adapter to the DVM input.

2Press Default Setup to set the oscilloscope to default conditions.

3 Select Acquisition from the Setup menu.

4Select Equivalent Time under Sampling Mode. Enable Averaging, and set the # Points for averaging to 32. Click Close.

Setup is the same as that for the voltage measurement accuracy test. See figure 3-5.

5Use the following table for steps 6 through 12.

Volts/div

Position

Supply

Tolerance

 

Limits

 

 

 

 

minimum

maximum

 

 

 

 

 

 

200 mV

2.00000 V

2.00 V

±36 mV

1.964 V

2.036 V

100 mV

1.00000 V

1.00 V

±18 mV

0.982 V

1.018 V

50 mV

500.000 mV

500 mV

±9 mV

491 mV

509 mV

 

 

 

 

 

 

6Select Channel 1 from the Setup menu.

7Set the vertical Scale to the Volts/div value from the first row of the table. Set the Offset to the Position value from the first row of the table. Click Close.

3–11

Page 48
Image 48