Agilent Technologies 54835A, 45A, 46A manual To reset the attenuator contact counter

Models: 54835A 46A 45A

1 207
Download 207 pages 4.15 Kb
Page 167
Image 167
To reset the attenuator contact counter

Chapter 6: Replacing Assemblies

To reset the attenuator contact counter

To reset the attenuator contact counter

1Turn off the Infiniium oscilloscope.

2Connect a mouse and a keyboard to the oscilloscope.

3Insert the service disk in the floppy drive and turn on the oscilloscope.

For units equipped with a 120 Mbyte floppy drive, do not insert the service disk until the message:

Starting Windows 9X...

appears on the screen.

4The oscilloscope will boot into the Windows® operating system.

5Press Control-Esc on the keyboard to access the Windows® Start menu.

6 Open the Programs group and click the Infiniium Scope icon.

7When the oscilloscope application comes up, click the mouse icon in the upper right corner of the screen.

8Click the Utility pull down menu, then click Self Test. In the Self Test screen click Select Service Extensions.

9Click the Self Test pull down menu, scroll down and select Atten Set.

Figure 6-28

Self Test

6–30

Page 167
Image 167
Agilent Technologies 54835A, 45A, 46A manual To reset the attenuator contact counter