Chapter 4: Calibrating and Adjusting

To check the power supply

To check the power supply

There are no adjustments for the supply. Perform this procedure only if you suspect a power supply problem.

Equipment Required

 

Equipment

Critical Specifications

Recommended Model/Part

 

Digital Voltmeter

Accuracy ±0.05%

Agilent 34401A

 

Procedure

 

 

1

Disconnect the oscilloscope power cord and remove the cover.

 

If necessary, refer to the procedures in chapter 6, “Replacing Assemblies.”

 

 

 

WARNING

INJURY CAN RESULT!

 

 

Use care when working around the oscilloscope with the cover off and power applied. The fan

 

blades are exposed and could cause injury; also, there is approximately 900 V generated by the

 

backlight inverter for the flat-panel display.

 

 

Connect the oscilloscope power cord and set power switch to ON.

2

3

Refer to the following figure for testpoint locations.

 

Figure 4-1

Power Supply Test Points

4–4

Page 69
Image 69
Agilent Technologies 45A, 46A manual To check the power supply, Disconnect the oscilloscope power cord and remove the cover