Agilent Technologies 46A, 45A, 54835A manual Replacing Assemblies, ESD Precautions, Tools Required

Models: 54835A 46A 45A

1 207
Download 207 pages 4.15 Kb
Page 139
Image 139
Replacing Assemblies

C A U T I O N

C A U T I O N

W A R N I N G

Replacing Assemblies

Use the procedures in this chapter when removing and replacing assemblies and parts in the Agilent Technologies 54835A/45A oscilloscope.

In general, the procedures that follow are placed in the order to be used to remove a particular assembly. The procedures listed first are for assemblies that must be removed first.

ESD Precautions

When using any of the procedures in this chapter you must use proper ESD precautions. As a minimum you must place the instrument on a properly grounded ESD mat and wear a properly grounded ESD wrist strap.

AVOID DAMAGE TO THE INSTRUMENT!

Failure to implement proper antistatic measures may result in damage to the instrument.

Tools Required

The following tools are required for these procedures.

Torx drivers: T8, T10, T15

Socket wrench: 5/8 inch

Medium size (3/16-in) flat-blade screwdriver

Nut Drivers: 3/16-in, 9/32-in, 5/16-in, 5/8-in

Torque driver, 0.34 Nm (3 in-lbs), 5 mm or 3/16-in hex drive

Torque driver, 0.34 Nm (3 in-lbs), Torx T6 drive

REMOVE POWER BEFORE REMOVING OR REPLACING ASSEMBLIES!

Do not remove or replace any circuit board assemblies in this instrument while power is applied. The assemblies contain components which may be damaged if the assembly is removed or replaced while power is connected to the instrument.

SHOCK HAZARD!

To avoid electrical shock, adhere closely to the following procedures. Also, after disconnecting the power cable, wait at least three minutes for the capacitors on the power supply and sweep boards to discharge before servicing this instrument. Hazardous voltages exist on the inverter for the display monitor.

6–2

Page 139
Image 139
Agilent Technologies 46A, 45A, 54835A manual Replacing Assemblies, ESD Precautions, Tools Required