Chapter 6: Replacing Assemblies

To reset the attenuator contact counter

10Click the Start Test button. A screen like the following will be displayed.

Figure 6-29

Attenuator Relay Actuations Setup

11In the Attenuator Channel box, select the channel you are changing.

12Under Set Value, enter the corrected attenuator value using the keyboard or the drop down keypad, then press the button labeled Apply ‘Set Value’ to Actuation Count.

13Repeat for each channel you need to change.

6–31

Page 168
Image 168
Agilent Technologies 45A Attenuator Channel box, select the channel you are changing, Attenuator Relay Actuations Setup