Offset Accuracy*2

± (1.00% of channel offset + 1% of full scale) at full-resolution channel scale.

dc Voltage Measurement

Dual Cursor

±[(dc gain accuracy)+(resolution)]

Accuracy*2

 

 

 

Single Cursor

±[(dc gain accuracy)

 

 

+(offset accuracy)+(resolution/2)]

AutoProbe Interface

AutoProbe is an intelligent communication and power link between compatible

 

probes and Infiniium scopes. AutoProbe completely configures the scope for the

 

attached probe. For instance, it identifies the probe type and sets up the proper input

 

impedance, attenuation ratio, probe power and offset range, as needed.

Horizontal

Main Time Base Range

100 ps/div to 20 s/div

 

Horizontal Position Range

pre-trigger

0 to -1 s or one full screen width,

 

 

whichever is larger.

 

post-trigger

0 to 1 s or one full screen width,

 

 

whichever is larger.

Delayed Sweep Range

1 ps/div to current main time base setting.

Delayed Sweep Delay Range

Within main time base acquisition record.

Resolution

2 ps

 

Timebase Accuracy

70 ppm (.007%)

 

t Accuracy*

Real Time mode4

±[(.007%)(t)+(0.2)(sample period)]

 

Equivalent Time mode

±[(.007%)(t)+(full scale/

 

 

(2 memory depth)) + 30 ps]

 

 

(Example: for 16 avgs, 9 ns signal,

1 ns/div, 1 channel, then accuracy =

±[(.007%)(9 ns)+(10 ns/(2 65,536)) + 30 ps] = 31 ps)