Chapter 5: Troubleshooting

No Display Trouble Isolation

No Display Trouble Isolation

This trouble isolation procedure helps isolate a problem to the assembly level when there is no display on the flat-panel liquid crystal display screen.

ARemove the cabinet and install the fan safety shield.

1 Disconnect the power cord from the oscilloscope. Refer to chapter 6, “Replacing Assemblies,” for instructions on removing the cabinet. Use care in handling the instrument.

2Install the optional fan safety shield, Agilent Technologies P/N 54801-00601, over the fans on the left side of the instrument.

WARNING

 

SHOCK HAZARD!

 

 

The backlight inverter assembly, which is mounted at the front corner of the instrument near

 

 

the flat-panel display, operates at 900 V at turn on. DO NOT handle this assembly while it is in

 

 

operation.

 

 

 

 

 

 

WARNING

 

INJURY CAN RESULT!

 

 

Once the cover is removed, the fan blades are exposed both inside and outside the chassis.

 

 

Disconnect the power cable before working around the fan. Use extreme caution in working with

 

 

the instrument when the cover is removed. Install the fan safety shield (Agilent Technologies

 

 

P/N 54801-00601) on the side of the chassis over the fan. Failure to observe these precautions

 

 

may result in injury.

 

 

3 Connect the power cord and press the power switch on the front panel.

 

 

 

B Check the fan connections.

 

 

Verify that the fan mounted in the left side of the instrument and the CPU fan (on the

 

 

motherboard) are connected.

 

 

1 The instrument fans connect to the side of the Acquisition assembly at J3 and J5.

 

 

2 The CPU fan connects to the top side of the motherboard at J46 (Fan Power).

 

 

3 Connect the power cord and press the power switch on the front panel. Verify that both

 

 

fans are running and that the front panel power switch LED is illuminated. These are

 

 

indications that the power supply is functioning. If the fans and LED are off, go to the

 

 

Power Supply Trouble Isolation flowchart.

 

C Connect an external monitor, cycle power, and observe the instrument power-on

 

 

sequence.

 

 

See table 5-1 to check the visual and audible power-on sequence, and repair as needed, referring

 

 

to other procedures as directed.

5–9

Page 86
Image 86
Agilent Technologies 54835A No Display Trouble Isolation, Remove the cabinet and install the fan safety shield, Sequence