To remove and replace the fan

Chapter 6: Replacing Assemblies

To remove and replace the fan

 

To remove and replace the fan

 

 

WARNING

AVOID INJURY!

 

The fan blades are exposed both inside and outside the chassis. Disconnect the power cable

 

before working around the fan. Use extreme caution in working with the instrument. Failure to

 

observe these precautions may result in injury.

 

 

Use this procedure to remove and replace the fan. When necessary, refer to other removal procedures.

1Disconnect the power cable and remove the cover.

2Disconnect the fan cable W21 from the acquisition board.

3 Remove the four fasteners securing the fan to the chassis.

To remove a fastener, unscrew the center pin using a #1 Posidrive screwdriver. You may need to apply upward pressure at the point of the pin with a small screwdriver. After removing the center pin, compress the fastener body and push it up and out of the fan and chassis. Be sure to support the fan before removing the last fastener.

Alternatively, you can cut the fastener at the outside of the chassis with a pair of diagonal cutters, then install new fasteners when you replace the fan.

Figure 6-24

 

 

Removing Fan Fasteners

 

 

 

CAUTION

 

AVOID OVERHEATING THE INSTRUMENT!

 

 

When replacing the fan, be sure the direction of the fan air flow is coming from the outside to

 

 

the inside of the instrument. Check the flow arrows on the fan and check for proper flow once

 

 

power is applied to the instrument. Improper air flow can overheat the instrument.

 

 

 

 

4 To install the fan, reverse this procedure.