Chapter 5: Troubleshooting

To configure the motherboard jumpers and set up the BIOS

Table 5-7

Series 727 System Configuration Settings

 

 

 

 

 

Configuration Item

Setting

 

Load Optimal Defaults

YES

 

Peripheral Setup (OnBoard Primary/Secondary IDE)

Primary

 

Detect IDE Utility (Hard drive Primary Master)

Data

 

Standard Setup—Date/Time

Enter current values

 

Standard Setup—Floppy Drive A

1.44 MB, 3-1/2

 

Advanced Setup—System Keyboard

Absent

 

Advanced Setup—External Cache

Disabled

 

Power Management/APM*

Enabled

 

Hard Disk Power Down Mode*

Suspend

 

Hard Disk Time Out (minute)*

15 min

* These configurations are only for instruments with 64 Mb or more of RAM.

7Exit the Setup program, saving the changes you made.

The oscilloscope should now reboot and come up with the scope display. If it does not, recheck your steps in the above procedure. If all the settings are correct, return to the Primary Troubleshooting Procedure.

5–36

Page 113
Image 113
Agilent Technologies 54835A, 45A Exit the Setup program, saving the changes you made, Configuration Item Setting, Primary