Unit Status Settings

Setting

Description

 

 

Asset Tag

An optional user settable field which is set via the RS-232 port. This could be a serial num-

 

ber, name or any other piece of information desired

 

 

Backlight Status

Shows OK if the electronics module detects that both banks of backlights are functioning

 

normally. Shows Failed if one or both of the backlight banks is not drawing the correct

 

amount of current

 

 

Internal Temperature

The current internal temperature at one point on the electronic module. This is not neces-

 

sarily the highest nor the lowest temperature point of the unit, nor is it an indicator of ambi-

 

ent temperatures.

 

If thermal guidelines (see “Normal Use Thermal Guidelines” on page 116) are exceeded

 

and additional ventilation or cooling has been implemented, this temperature may be used

 

as a datapoint to determine if the temperature of the unit is being affected.

 

 

Mode ID

The index into the internal mode table found for the incoming signal. This is used in diagno-

 

sis of mode detection problems

 

 

HPer

The measured horizontal period of the incoming signal. This is used in diagnosis of mode

 

detection problems

 

 

VLines

The number of horizontal lines detected per vertical period. This is used in diagnosis of

 

mode detection problems

 

 