11.4Diagnostic Messages, Continued

Diagnostic Messages,

Table 33

Communication Status Message Table

continued

 

 

SMV Status

SCT Status Message SFC Display Message TDC Status Message

Possible Cause

What to Do

-

-

-

Command Aborted

Communication Error

Upload failed

Download Failed

TAG NO.

COMM ABORTED

TAG NO.

END AROUND ERR

SAVE/RESTORE RESTORE FAILED

-

-

-

Communications aborted.

SFC – Pressed [CLR] key during communications operation.

Communications unsuccessful.

Database restore or download function failed due to a problem with the current configuration or a communications error.

Retry aborted operation.

Check loop wiring and STC/SFC connections.

If error persists, replace transmitter electronics module.

Check transmitter and try again.

-

-

Invalid Response

Illegal operation

TAG NO.

ILLEGAL RESPONSE

URV 3 . TAG ID.

INVALID REQUEST

-

-

The transmitter did not respond properly since the response was not recognizable. The message was probably corrupted by external influences.

Transmitter sent illegal response to SCT or SFC.

Requesting transmitter to correct or set its URV to a value that results in too small a span, or correct its LRV or URV while in input or output mode.

SFC – Keystroke is not valid for given transmitter.

SCT – The requested transaction is not supported by the transmitter.

Try communicating again.

Check that correct URV calibration pressure is being applied to transmitter, or that transmitter is not in input or output mode.

Check that keystroke is applicable for given transmitter.

Make sure the device version is compatible with the current release of the SCT 3000.