Honeywell user manual SMV 3000 Transmitter User’s Manual 133

Models: SMV 3000

1 217
Download 217 pages 57.98 Kb
Page 145
Image 145

11.4 Diagnostic Messages, Continued

Diagnostic Messages,

Table 33

Communication Status Message Table, continued

continued

 

 

SMV Status

-

-

-

SCT Status Message

SFC Display Message

TDC Status Message

 

 

 

 

 

-

STATUS

TAG

ID.

-

 

NACK

RESPONSE

 

 

 

 

 

 

-

TAG NO.

 

 

-

 

FAILED

COMM

CHK

 

 

 

 

 

 

-

TAG

NO.

-

 

HI

RES/LO VOLT

 

 

 

 

 

Possible Cause

Transmitter sent a negative response because it could not process one or more commands.

SFC failed a communications diagnostic check. Could be an SFC electronic problem or a faulty or dead communication loop.

Either there is too much resistance in loop (open circuit), voltage is too low, or both.

What to Do

Check configuration and try again.

• Check polarity and try again.

• Press [stat] key and do any

 

corrective action required and try

 

again.

Check communication loop.

Replace SFC.

• Check polarity, wiring, and

 

power supply. There must be 11

 

volts minimum at transmitter to

 

permit operation.

• Check for defective or

 

misapplied capacitive or

 

inductive devices (filters).

-

-

TAG NO.

 

-

No response from transmitter.

 

 

NO XMTR

RESPONSE

 

Could be transmitter or loop

 

 

 

failure.

 

 

 

 

 

 

 

 

 

 

 

• Try communicating again.

• Check that transmitter’s loop

integrity has been maintained,

that SCT or SFC is connected

properly, and that loop

resistance is at least 250.

SCT – Select Tag ID from the View pull down menu.

SFC – Press [ID] key and do any corrective action required and try again.

1/99