6.9FlowConf Configuration - PV4, Continued

About LRV and URV

 

The LRV and URV set the desired zero and span points for your

 

 

calculated measurement range as shown in the example in Figure 24.

 

Figure 24

Typical Volumetric Flow Range Setting Values

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LRL

Typical Range Configuration for Volumetric Flow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LRV

 

 

 

SPAN

 

 

URV

 

 

 

 

URL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

m3/h

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1300

 

 

 

 

0

 

 

 

325

 

 

650

975

 

 

 

 

 

Range Limits

 

Measurement

 

Lower Range

 

Upper Range

 

Span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Range

 

Value

 

Value

 

 

 

 

 

 

 

 

 

0 to 1300 m3/h

 

 

0 to 650 m3/h

 

 

0 m3/h

 

650 m3/h

 

650 m3/h

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ATTENTION

The default engineering units for volumetric flow rate is cubic meters per hour and tonnes per hour is the default engineering units for mass flow rate.

The URV changes automatically to compensate for any changes in the LRV and maintain the present span (URV – LRV).

If you must change both the LRV and URV, always change the LRV first.

Continued on next page