Honeywell SMV 3000 user manual SCT 3000 On-line Connections to, Step Action

Models: SMV 3000

1 217
Download 217 pages 57.98 Kb
Page 51
Image 51

5.2 Establishing Communications, Continued

ATTENTION

Connecting the host computer to an SMV for on-line communications requires Smartline Option Module consisting of a PC Card and Line Interface Module.

SCT 3000 On-line Connections to the

SMV

WARNING

Table 8 provides the steps to connect the assembled SCT 3000 hardware between the host computer and the SMV for on-line communications.

When the transmitter’s end-cap is removed, the housing is not explosionproof.

Table 8 Making SCT 3000 Hardware Connections

Step

Action

 

 

1With the power to the host computer turned off, insert the PC Card into the type II PCMCIA slot on the host computer (see Figure 5-1).

ATTENTION To use the SCT 3000 in a desktop computer without a PCMCIA slot, you must install a user-supplied PCMCIA host adapter. Honeywell has performance- qualified the following PCMCIA host adapters for use with the SCT:

--TMB-240 Single Slot Internal Front Panel Adapter

--TMB-250 Dual Slot Internal Front Panel Adapter

--GS-120 Greystone Peripherals, Inc.

--GS-320 Greystone Peripherals, Inc.

CAUTION Do not insert a PC Card into a host computer’s PCMCIA slot while the host computer is powered on.

2Remove the end-cap at the terminal block side of the SMV and connect the easy hooks or alligator clips at the end of the adapter cable to the respective terminals on the SMV as follows:

Connect the red lead to the positive terminal.

Connect the black lead to the negative terminal.

ATTENTION The SCT 3000 can be connected to only one SMV at a time.

Continued on next page