7.3 Running Output Check, Continued

Procedure

Table 20

Step

1

2

3

4

5

Output Check for SMV Transmitters in DE Mode

Action

Connect SCT to SMV and establish communications. (See Subsection 5.2 for procedure, if necessary.)

Be sure any switches that may trip alarms or interlocks associated with analog loops are secured or turned off.

Perform Upload of the SMV database to the SCT.

Select General tab card and set communication mode to Digital Enhanced.

Set any of the SMV transmitter PVs to output mode, by selecting the appropriate tab cards.

DPOutCal, (for PV1)

APOutCal, (for PV2)

TempOutCal, (for PV3) or

FlowOutCal, (for PV4)

6

7

8

9

Enter an output value and then set PV to Output mode.

Open the PV Monitor window by selecting PV Monitor from the View pull down menu. Read the PV outputs.

Also, check the PV outputs as displayed at the digital receiver.

Select appropriate tab card for the PVs that were set to output mode and clear the output mode.

Select Status tab card to verify that all transmitter outputs are in not in output mode and that there are no new messages.