8.2Accessing Operation Data, Continued

Procedure, continued

Table 25 Accessing Transmitter Operation Data Using SCT,

Continued

IF you want to view…

1.the input value for a given PV, which is updated every six seconds.

2.the present transmitter output in percent for a given PV, which is updated every six seconds.

Select the SCT

Window or Tab Card

PV Monitor Window

And . . .

Read: PV Input

PV % of span

1.the span, which is the URV minus the LRV for a given PV.

2.the Upper Range Limit of a given PV.

3.the Lower Range Limit of a given PV.

DPConf (for PV1)

APConf (for PV2)

TempConf (for PV3)

FlowConf (for PV4)

Read: Span

URL

LRL

the failsafe output direction for the transmitter.

ATTENTION You can change the default failsafe direction from upscale to downscale. See Section 8.3, “Changing Default Failsafe Direction”.

the present meter body temperature (±5 °C) measured by circuitry in the transmitter’s sensor.

ATTENTION You can change the temperature engineering units to °F, °R or °K by selecting the SV Units field in the DPConf tab card.

General Tab Card

PV Monitor Window

Read:

Analog Failsafe

Direction

Click on SV button on DP gauge

Read: SV

Continued on next page

1/99

SMV 3000 Transmitter User’s Manual

93