9. Troubleshooting

Problem

Cause/Remedy

 

 

Fluorescence

 

 

 

The image is completely dark (no fluorescence).

￿Open the shutter (p. 67).

 

￿Select the incident-light axis (IL) (p. 65).

 

￿Check your specimen, e.g. its antibody bind-

 

ing.

 

￿Insert a new lamp (p. 45ff).

The fluorescence is too weak.

LeicaScreen

￿Insert the booster (p. 90). ￿Center the lamp (p. 77ff) ￿Insert a new lamp (p. 45ff).

Init Error!

￿Check the cable connections.

 

￿Check whether the cover of the filter disk has

 

clicked into position.

 

￿Check the installed objectives, filter cubes,

 

etc.

 

￿Switch the microscope off and back on.