6. Assembly

6.2Installation of the

transmitted-light illumination carrier (DL)

Wipe the installation surface on the microscope (8.3) with a dry cloth. Tip the illumination carrier

(8.1) back slightly and install it so that the pin

(8.2) engages the groove in the support surface

(8.4).

Set the DL illumination carrier upright and fasten it with the 4 screws.

When fastening the transmitted-light illumina- tion carrier, do not hold it so as to ensure its op- timal alignment with the optical axis.

The tilt angle of the illumination carrier can be varied with the knurled screw (9.1) or fixed verti- cally.

Fig. 8 Installing the transmitted-light illumination carrier

1Transmitted-light illumination carrier

2Transmitted-light illumination carrier pin

3Support surface

4Support surface groove

5Support surface groove

6EXT1-EXT4 sockets

7Electronics cable

1

3

7

4

Leica DMI4000 B and Leica DMI6000 B

Connect the electronics cable to one of the sockets, EXT1 - EXT4.

The transmitted-light lamp housing for 12 V 100 W halogen lamps is a separate component. For instructions on replacing the halogen lamp

Ch. 6.10, p. 45.

Fig. 9 Transmitted-light illumination carrier, rear side

1Knurled locking knob

of the transmitted-light illumination carrier

2Connector cable for the electronics box

1

2

1

 

 

1

 

 

 

 

5

 

 

5

 

 

2

6

30