7. Start-up

Function buttons on the front panel (Fig. 87)

100% of the light goes to the eye- piece (87.1).

Toggle function

for

the side ports

 

(87.2). This function depends on the

 

 

 

 

individual microscope configuration.

 

 

Note:

 

 

 

 

Switching to the bottom port:

 

 

via the variable function buttons

 

 

(Leica DMI6000 B only), switching to

 

 

top port: manually.

 

SHUTTER

Opens and closes the shutter (87.3).

 

 

Switches between the possible mag-

nifications of

the

magnification

 

 

changer (87.4).

 

 

The magnification changer is set to the

1x

magnification 1x (87.5).

Fig. 87 Front control panel

1100% light to eyepiece

2Toggle ports

3Shutter

4Switch between subsequent magnifications

5Subsequent magnification 1x

6Selecting filter cubes

1 2 3 4 5

6

6

CUBE The CUBE 1 to CUBE 6 (87.6) buttons permit the direct selection of indi- vidual filter cubes, provided the se- lected cube is valid for the selected method.

Press the CUBE 3 and CUBE 4 but- tons at the same time to display the assignments of the variable function buttons. To reset the display, press the buttons again or wait 3 seconds.

Focus buttons (Fig. 88) (DMI6000 B only)

Z

Moves the Z drive in the indicated di-

Z

rection.

SET + ZSets the upper focus stop.

SET + ZSets the lower stop.

Fig. 88

1Focus control buttons

2Open filter drawer

2

1

67

Page 67
Image 67