6. Assembly

6.10Installation and replacement of the transmitted-light lamps: 107 or 107/2 lamp housing

This lamp housing is used with a 12V 100W halo- gen lamp, which is already mounted.

In case the lamp has to be removed:

Changing the 12 V 100 W halogen lamp

Caution!

Ensure that the lamp housing has been dis- connected from the power supply. Unplug the power plug and the power supply during assembly.

Caution!

Light sources pose a potential irradiation risk (glare, UV-radiation, IR-radiation). Therefore, lamps have to be operated in closed housings.

¥Remove the fastener screw on the housing (Fig. 50a).

¥Lift the housing off (Fig. 50b).

¥Remove the lamp.

Caution!

Do not remove the new lampÕs dust cover until you have installed the lamp. Avoid fin- gerprints on the lamp.

¥Insert the new 12 V 100 W lamp (Fig. 51) with the dust cover straight into the socket until it stops. Be sure that the lamp is inserted straight.

¥Remove the lampÕs dust cover.

¥Replace the housing and fasten it in place us- ing the fastening screw.

Fig. 50b

Removing housing

Fig. 50a

Fig. 50c

 

Lamp housing 107/2

Lamp housing 107/2

1

Releasing the

opened

fastening screw

1

Mount with

 

 

 

halogen lamp

 

 

2

Collector

 

 

 

 

2

45