9. Troubleshooting

Problem

Cause/Remedy

 

 

Phase contrast

No phase contrast is possible.

￿The specimen is too thick.

￿The cover glass is not placed evenly. ￿Check the centering of the light rings

(p. 73).

Polarization

No polarization contrast is possible.

￿Bring the polarizer and analyzer into cross po- sition until they reach maximum darkness (without specimen).

(p. 86).

Transmitted light interference contrast

No transmitted light interference contrast is

￿The specimen is too thick or too thin.

possible

￿Embedding medium or specimen are of bire-

 

fringent material. Rotate the specimen.

 

￿The difference in the refractive indices of the

 

specimen and the embedding medium is too

 

small.

 

￿The cover glass is too thick.

 

￿Check the Koehler illumination (p. 69).

 

￿Bring the polarizer and analyzer into cross po-

 

sition until they reach maximum darkness

 

(without specimen).

 

(p. 87).

 

￿Check whether the suitable condenser prism

 

and corresponding objective prism are se-

 

lected (manual alternative p. 87).

 

￿Ensure that the IC prisms are correctly seated

 

(p. 40).

104

Page 104
Image 104
Leica DMI4000B Troubleshooting Problem Cause/Remedy Phase contrast, Polarization, Transmitted light interference contrast