6. Assembly

6.8 Installation of filters in the illumination arm

The Leica DMI series is equipped with a filter magazine to accommodate two 40 mm dia. filters as a standard feature. The filters are installed at the factory. To change filters yourself, proceed as follows:

¥Release the screw (46.1) and remove the cover.

¥Place the filter in the holder.

¥Place the cover on transmitted-light illumination

carrier and fasten with the locking screw. Leica DMI6000B:

¥ Activate the filters via the Leica Application Suite (LAS).

Leica DMI3000B and Leica DMI4000 B:

¥Mark the 2 levers with the provided adhesive la- bels.

Fig. 46 Unscrewing the filter holder cover and inserting filters in the transmitted-light illumination arm

1Screw

6.9Installing the transmitted-light lamp housing

¥Place the lamp housing in the transmitted light lamp housing mount (Fig. 47) and fasten it with the clamping screw on the side.

¥Thread the cable through the transmitted-light illumination arm (Fig. 48).

¥Connect the lamp housing cable to the power supply for transmitted light on the Leica

CTRxxxx electronics box (Fig. 49.1). Leica DMI3000 B:

¥For the DMI3000 B, connect the cable directly to the back of the microscope.

For instructions on changing the lamp, please see Chapter 6.10.

These instructions also apply to installing an Hg lamp on the transmitted-light axis. For descrip- tions of the lamp housings and replacement of the burner, please see Chapter 6.12, p. 48ff.

Fig. 47 Mounting the lamp housing on the transmitted-light illumination arm

1

Fig. 48 Lamp housing cabling (cable duct)

Fig. 49 Connecting the lamp housing to the Leica CTR6000 electronics box

1

44

Page 44
Image 44