6. Assembly

Caution!Caution!

Make sure to follow the safety notes on page 48.

¥To open the 106 z lamp housing, unscrew the fastening screws on the cover l. Loosen the contact plug somewhat and pull it out of the socket (63.9). Flip the cover up (63.1).

¥Loosen the mounting screws (63.8) on the lamp socket and pull the socket out.

¥Remove the transport anchorage (red plastic rod in place of the burner) in the lamp mount. To do so, remove the lower clamp (60.1, 61.1). Pull up the cooling element (61.3, 60.3) and turn it to the side. Detach the lower clamp system (61.2, 60.2) and remove the transport anchorage.

Do not remove the burnerÕs dust cover until you have installed the lamp. Avoid fingerprints on the lamp. Sweat from your fingers on the glass will shorten the life of the lamp significantly.

¥ Install the burner in reverse order.

Caution!

Xe 75 burner:

Remove the burnerÕs dust cover (61.4) after you have installed the burner.

Fig. 63 106 z lamp housing (on the side, open)

1Cover raised

2Collector

312V 100W lamp or

gas discharge lamp in mount

4Reflector (mirror)

5, 6, 7 Adjusting screw for x-y reflector

8Locking screws for lamp mount

9Socket for contact plug

Fig. 62 Rear panel of ebq 100 supply unit

1

1 Lamp connection

 

1

2

3

4

5

6

7

8 9 8

50

Page 50
Image 50