6. Assembly

6.13 Equipping the Incident Light Turret Disc

!Caution:

Please read this section completely before be- ginning with the assembly of the turret disk. Leica DMI4000 B and Leica DMI6000 B:

The fluorescence drawer is located on the right side of the stand. Before opening this drawer, remove the cap below the drawer covering the analyzer slot (66.1). Remove the analyzer if it is already in the slot.

The replacement of individual cubes is more convenient with the microscope switched on. The position to be changed then automatically turns to the outside and you can be sure that the cube is positioned in the correct holder. You can therefore postpone installing the filter cubes un- til after the microscope has been switched on.

You can also insert the filter cubes while the in- strument is switched off.

Press the white button next to the drawer. The drawer will glide out into its initial position.

The inner disk can only be turned to accept the fluo cubes in this position.

Fig. 66 Opening the fluorescence drawer 1 Analyzer slot

1

Fig. 67 Open fluorescence drawer

Fig. 68 Inserting or removing a filter cube

Fig. 64 Filter cube,

Fig. 65 Filter cube,

front side

back side