TONER FULL state
240 ms < t1 < 1.5 sec
t1
4.875 sec.
TONER LOW state
|
| t1 > 1.5 sec. |
or | ||
| t1 | t1 < 240 ms |
| 4.875 sec. |
|
•When the Toner Low state is detected 2 times consecutively, Toner Low is established.
•When the Toner Full state is detected 2 times consecutively, Toner Low is cancelled.
•When there is no change with the toner sensor for 2 cycles (4.875 sec. x 2) or more, then the Toner Sensor Alarm is activated.
•The toner sensor is not monitored while the drum motor is in halt.