Teledyne 300 instruction manual Specifications

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 10
Image 10
1.4 Specifications

1.4 Specifications

Ranges

User selectable to any full-scale range from 1 ppm

 

to 1,000 ppm

Zero Noise

< 0.025 ppm (rms)

Span Noise

< 0.5% of reading (rms)

Lower Detectable Limit

< 0.050 ppm

Zero Drift (24 hours)

* <0.1 ppm

Zero Drift (7 days)

* <0.2 ppm

Span Drift (7 days)

* 1% of reading

Linearity

1% FS

Precision

0.5% of reading

Lag Time

10 sec

Rise/Fall Time (95%)

<60 sec

Sample Flow Rate

800cc/min. ± 10%

Temperature Range

5-40°C

Humidity Range

0-95% RH, non-condensing

Temp Coefficient

< 0.05 % per °C

Voltage Coefficient

< 0.05 % per V

Dimensions HxWxD

7"x 17"x 25"

 

(178mm x 432mm x 660mm)

Weight

50 lbs (22.7 kg)

Power

110V~/60Hz, 220V~/50Hz., 240V~/50Hz. 250 Watts

Power, CE

230V~/50Hz, 2.5 A

Environmental Conditions

Installation Category (Overvoltage Category) II

Recorder Outputs

Pollution Degree 2

± 100 mV, ± 1 V, ± 5 V, ±10 V (Bi-Polar)

Status

12 status outputs from opto-isolators

* at constant temperature and voltage

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 4

Page 10
Image 10
Teledyne 300 instruction manual Specifications

FAQ

What are the temperature and humidity ranges for the Teledyne API Model 300 CO Analyzer?

The temperature range is 5-40C and the humidity range is 0-95% RH, non-condensing.

What is the power requirement for the Teledyne API Model 300 CO Analyzer?

The power requirement is 110V~/60Hz, 220V~/50Hz, 240V~/50Hz, 250 Watts.

What are the recorder outputs available for the Teledyne API Model 300 CO Analyzer?

The recorder outputs include pollution degree 2, 100 mV, 1 V, 5 V, 10 V Bi-Polar, and 12 status outputs from opto-isolators.