There are two acceptable methods of generating accurate CO concentrations for calibrating the Model 300.
One method uses a single cylinder of
Alternatively, several cylinders of
The cylinder concentrations must be traceable to NIST standards.
Flow correction for standard temperature and pressure (STP) is not required with either method. With dilution, the correction is
8.2.1PROCEDURE
1.Set the Analyzer to the desired range. (The
2.Set the calibration system to deliver a flow of at least 1500 cc/min. The Model 300 draws approximately 800 cc/min. (See Section 8.1 and Figure 8.2 for vent flow calculation.)
3.
4.Connect the Analyzer REC (recorder) terminals to a calibrated
The standard output voltage of
If, in service, data is to be collected from a device (printer) connected to the
5.Set the calibrator to deliver zero air to the manifold.
6.Push “CAL” on the Analyzer front panel.
7.Enter password. (If Enabled)
8.Wait 15 minutes for the Analyzer to stabilize.
9.Push “ZERO” and “ENTR” on the front panel. The Analyzer is now “zeroed.”
10.Push “EXIT,” “EXIT.” (Return to sample mode.)
11.Record the DVM reading and the percentage chart reading or the
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 74