9.4 Output voltage range changes

To view the current dark offset, press SETUP-MORE-DARK-VIEW.Press EXIT when finished. No password is required to view the dark offset, only to change it.

9.4 Output voltage range changes

Output voltage ranges are set by DIP Switch settings on the V/F board. To change the range for the analog outputs:

1.Turn off instrument power. Remove the instrument cover. Locate the V/F board near the top of the drawing using Figure 1.6.

2.Locate switches S1, S2, and S3 along the top edge of the card. Select the desired range by setting the switches as shown in Table 9.1, below.

V/F BOARD SWITCH SETTINGS - RANGES FOR ANALOG OUTPUT

Switch

100 mV

1 V

5 V

10 V

 

Full Scale

Full Scale

Full Scale

Full Scale

S1 (Recorder Output)

1, 6

1, 5

1, 4

1, 3

S2 (DAS Output)

1, 6

1, 5

1, 4

1, 3

S3 (Test Output)

1, 6, 7

1, 5, 7

1, 4, 7

1, 3, 7

TABLE 9.1

NOTE

TO ADJUST ANALOG RECORDER OFFSET,

SEE SECTION 4.7.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 80

Page 86
Image 86
Teledyne 300 Output voltage range changes, Switch, 100 mV, Full Scale, To Adjust Analog Recorder Offset See Section

FAQ

How do I view the current dark offset?

To view the current dark offset, press SETUP-MORE-DARK-VIEW and press EXIT when finished.

How can I change the output voltage ranges for the analog outputs?

Output voltage ranges are set by DIP switch settings on the V/F board. Follow the instructions in the manual to locate and change the desired range.

Is a password required to change the dark offset?

Yes, a password is required to change the dark offset. Refer to the manual for more information on setting and changing the password.