THE PROGRAMMED START TIME MUST BE A MINIMUM OF 5 MINUTES LATER THAN THE REAL TIME CLOCK (See Section 4.3 and 4.4 for setting real time clock).

minute (00 - 59). Enter the time of day for calibration check and then press ENTR to accept the new time or EXIT to leave the time unchanged.

NOTE

THE PROGRAMMED START TIME MUST BE A MINIMUM OF 5 MINUTES LATER THAN THE REAL TIME CLOCK (See Section 4.3 and 4.4 for setting real time clock).

Delta Days: The number of delta days is the number of days between each auto- sequence. Enter desired number of delay days(0-365) and press ENTR.

Delta Time: The delta time allows the automatic Z/S check time-of-day to be delayed in the format of HH:MM, where HH is the hour from 00 to 23 and MM is the minutes(00- 59). The delta days and delta time are added together to determine the total delay between sequences. The delta time parameter allows you to advance or retard the starting time by a fixed amount each time the sequence is run. For example: Setting the delta days to 1 day and the delta time to 15 minutes will delay the starting time for the sequence by 15 minutes each day. If you want to have the sequence run at the same time every day, simply set the delta time to zero.

NOTE

Avoid setting two or more sequences at the same time of the day. Any new sequence which is initiated whether from a timer, the RS-232, or the contact closure inputs will override any sequence which is in progress.

Duration: The duration of each step of the sequence. Enter the duration in minutes(1-60) and press ENTR.

Calibrate: When set to ON, the sequence will adjust the internal formulas (slope and offset) in the Analyzer to the value set in the span variable. If this feature is enabled along with the automatic zero/span check, the Analyzer will re-adjust its formulas to match the predetermined zero and span settings once each day.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 34

Page 40
Image 40
Teledyne 300 instruction manual

FAQ

How do I calibrate the Teledyne API Model 300 CO Analyzer?

To calibrate the Analyzer, set the Calibrate feature to ON in the sequence settings. This will adjust the internal formulas slope and offset to match the predetermined zero and span settings once each day.

What is the minimum start time for calibration check?

The programmed start time must be a minimum of 5 minutes later than the real time clock. Refer to Section 4.3 and 4.4 for setting the real time clock.

How does the delta time affect the automatic Z/S check time-of-day?

The delta time allows you to delay the starting time by a fixed amount each time the sequence is run. The delta days and delta time are added together to determine the total delay between sequences.