BE CAREFUL USING THE BUBBLE

CAUTION

BE CAREFUL USING THE BUBBLE

SOLUTION. IF THERE IS NO INTERNAL

PRESSURE, THE SOLUTION MAY ENTER

AND CONTAMINATE THE CELL.

DO NOT ATTEMPT TO USE THE BUBBLE SOLUTION

WHILE THE UNIT IS UNDER VACUUM. THIS MAY

CAUSE DAMAGE TO THE ANALYZER.

USE ONLY BUBBLES, NOT LIQUID.

11.3.2 leak self test

If only a Model 300 is available, the following alternate method for leak checking can be used:

The power must be on and the pump must be in-line. Disconnect the power from the pump. Check for the sample flow. If the sample flow does not read zero, record the reading. Re-connect the power to the pump.

Cap the sample inlet. Check if the sample flow goes down to zero. If not, check the sample flow exceeds the prerecorded value when the pump was disconnected. If the sample flow is above zero or higher than the prerecorded value, the unit has the leak.

11.4 Changing the prom

1.Locate the CPU card by referring to Figure 1.6.

2.Remove the screws that hold the CPU card (SBC40 printed on the lever) top corner to the A/D - I/O card, then remove the card from the STD-BUS backplane.

3.Remove the two cables attached to the SBC40, taking note of the polarity.

4.Remove the card, laying it down on an insulating surface such that the card edge pins on the PCB are on the left. The PROM chip should be at the top center. The current chip should be labeled with something like "API CO A.6 - - -". Gently pry the chip from its socket and replace it with the new chip. Install the chip in the left end of the socket with the notch facing to the right. Make sure that all of the legs insert into the socket correctly.

5.Replace the CPU board and re-attach the connectors, making sure to observe the

polarity.

6 Re-attach the CPU card to the STD-BUS.

7.Move the power switch to the "ON" position and observe the front panel display. As the analyzer goes through the setup the version number will be displayed on

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 103

Page 109
Image 109
Teledyne 300 Changing the prom, leak self test, Be Careful Using The Bubble, Solution. If There Is No Internal

FAQ

How do I perform the leak self test?

If only a Model 300 is available, follow the alternate method for leak checking as outlined in the manual.

How do I change the prom?

Refer to the manual for step-by-step instructions on changing the prom of the Teledyne Model 300 Carbon Monoxide Alarm.

Can I use bubble solution for the alarm?

Take caution when using the bubble solution to avoid contamination of the unit. Refer to the manual for important guidelines.