1.8.3 Internal zero/span

The IZS option includes the valves and connection parts described above, and in addition, includes an internal zero air scrubber. This scrubber operates by catalyticly converting CO in the gas stream to CO2. If the Analyzer is equipped with the IZS option, Span gas connections should be made as described above and a source of clean air at ambient pressure should be connected to the IZS/Zero Air Port on the Analyzer's rear panel.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 21

Page 27
Image 27
Teledyne 300 instruction manual Internal zero/span

FAQ

What is included in the IZS option?

The IZS option includes valves, connection parts, and an internal zero air scrubber that catalytically converts CO in the gas stream to CO2.

How should span gas connections be made with the IZS option?

Span gas connections should be made as described in the manual, with a source of clean air at ambient pressure connected to the IZS/Zero Air Port on the Analyzer's rear panel.