the front panel. It should read the same as the version number that was located on the top right corner of the label on the PROM.

8.Re-enter any non-default settings such as RANGE or ASEQ. Re-enter the CONC value in the CAL menu. Check all settings to make sure that expected setup parameters are present.

9.Re-calibrate the analyzer so that the default slope and intercept are overwritten with the correct values.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 104

Page 110
Image 110
Teledyne 300 instruction manual

FAQ

How do I re-enter non-default settings?

To re-enter non-default settings, such as RANGE or ASEQ, go to the CAL menu and re-enter the CONC value.

What should I do if setup parameters are missing?

Check all settings to ensure that the expected setup parameters are present. If any are missing, re-enter them as needed.

Where can I find the version number on the front panel?

The version number on the front panel should match the version number located on the top right corner of the label on the PROM.