2.1.6 RS-232 interface

2.1.6 RS-232 interface

The Teledyne API CO Analyzer features an RS-232 interface which can output the instantaneous and/or average CO data to another computer. It can also be used as a command and status channel to allow another computer to control the Analyzer. Refer to Figure 1.3 for details on using the RS-232 interface. Tips on connecting the RS-232 port can be found in Appendix B.

2.1.7 Password protection

The Teledyne API CO Analyzer provides password protection of the calibration and setup functions to prevent incorrect adjustments to the Analyzer. There are three levels of passwords which correspond to operator, supervisor/maintenance, and analyzer configuration functions. When prompted for a password, any of the valid passwords can be entered, but the CPU will limit access to the functions allowed for that password level. Each level allows access to the functions of all the levels below plus some additional functions. Table 2.1 lists the password levels and the functions allowed for each level.

PASSWORD LEVELS

Password

Level

Functions Allowed

No password

0

TEST

 

 

MSG

 

 

CLR

Operator (512)

1

CALZ

 

 

CALS

 

 

CAL

Setup (818)

2

SETUP

 

 

SETUP-VARS

 

 

SETUP-DIAG

Supervisor

3

ALL

(101)

 

 

 

TABLE 2.1

 

NOTE

THE OPERATOR AND SETUP PASSWORDS CAN BE

DISABLED. TO DO THIS, ENTER SETUP-PASSWORDAND

TOGGLE “ON” TO “OFF”. PUSH “ENTR” AND THE

PASSWORDS WILL BE DISABLED

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 24

Page 30
Image 30
Teledyne 300 instruction manual 2.1.6 RS-232interface, Password protection, Password Levels, Functions Allowed

FAQ

What is the purpose of the RS-232 interface?

The RS-232 interface allows the CO Analyzer to output instantaneous and/or average CO data to another computer and can be used as a command and status channel for controlling the Analyzer.

How many levels of passwords are available for the Analyzer?

There are three levels of passwords for operator, supervisor/maintenance, and analyzer configuration functions, each allowing access to certain functions.

How to disable the operator and setup passwords?

To disable the operator and setup passwords, enter the setup-password and toggle on to off, then push enter to disable the passwords.