10.4 Troubleshooting using test function values

10.4 Troubleshooting using test function values

The Model 300 provides the capability to display, on operator demand, the values of Test Functions which allow the observation of key analyzer operating parameters. These Test Functions can be accessed by depressing the TEST Button on the instrument's front panel, with each depression of the button causing the next test function to be displayed. By comparing the values of Test Functions to acceptable operating limits, it is possible to quickly isolate and correct most problems.

NOTE

IF THE VALUE OF ANY TEST FUNCTION IS DISPLAYED

AS “XXXX”, THIS INDICATES THAT THE READING IS

OFF SCALE OR OTHERWISE NON-VALID.

Table 10-2 provides a list of available Test Functions along with their meaning, their range of acceptable values, and the recommended corrective actions if the value is not in the acceptable range. Additionally, Table 1-1 in this manual provides a list of the values of all Test Functions at the time the analyzer left the factory.

TEST FUNCTION VALUES

Test Function

Meaning

Acceptable Values

Corrective Action

 

 

 

for Unacceptable

 

 

 

Values

RANGE

The Current Full

Any

None required

 

Scale Range Setting

 

 

 

of the analyzer's

 

 

 

analog outputs

 

 

CO MEAS

The most recent

2000-4800 CNTS

Check and adjust IR

 

detector reading

 

source and Sync

 

taken in Measure

 

Demodulator and

 

mode

 

optical alignment as

 

 

 

described in Sect.

 

 

 

10.6.6 and 10.6.8

CO REF

The most recent

2000-4800 mV

Check and adjust IR

 

detector reading

 

source and Sync

 

taken in Reference

 

Demodulator and

 

mode

 

optical alignment as

 

 

 

described in Sect.

 

 

 

10.6.6 and 10.6.8

MR RATIO

The ratio of the CO

1.00-1.25

Check CO REF and

 

MEAS value to CO

 

CO MEAS values as

 

REF value

 

described above.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 87

Page 93
Image 93
Teledyne 300 Troubleshooting using test function values, If The Value Of Any Test Function Is Displayed, Acceptable Values

FAQ

What is the purpose of the test function values?

The test function values allow the observation of key analyzer operating parameters and help in isolating and correcting most problems.

How can I access the test function values?

The test function values can be accessed by depressing the TEST Button on the instruments front panel.

What should I do if a test function value is displayed as XXXX?

If the value of a test function is displayed as XXXX, it indicates that the reading is off scale or otherwise non-valid.