Teledyne 300 instruction manual St Flow Alarm

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 62
Image 62

13

ST_FLOW_ALARM

YES

Status Bit - Flow alarm

 

 

 

Logic High = Sample flow out of spec

 

 

 

Logic Low = Flows within spec

14

ST_TEMP_ALARM

YES

Status Bit - Temperature alarm

 

 

 

Logic High = One or more temps out

 

 

 

of spec

 

 

 

Logic Low = Temps within spec

15

ST_DIAG_MODE

YES

Status Bit - In Diagnostic mode

 

 

 

Logic High = M100A in Diagnostic

 

 

 

mode

 

 

 

Logic Low = Not in Diag mode

16

ST_POWER_OK

YES

Status Bit - Power OK

 

 

 

Logic High = Instrument power is on

 

 

 

Logic Low = Instrument power is off

17

ST_PRESS_ALARM

YES

Status Bit - Flow alarm

 

 

 

Logic High = Sample pressure out of

 

 

 

spec

 

 

 

Logic Low = pressure within spec

18

ST_HIGH_RANGE

YES

Status Bit - Autorange High Range

 

 

 

Logic High = M300 in high range

 

 

 

Logic Low = M300 in low range

19

ST_SYSTEM_OK

YES

Status Bit - System OK

 

 

 

Logic High = No instrument warning

 

 

 

present

 

 

 

Logic Low = 1 or more alarm present

20

ST_BENCH_ALARM

YES

Status Bit - Bench Temperature Alarm

 

 

 

Logic High = Bench Temp out of spec

 

 

 

Logic Low = Bench Temp in spec

21

ST_SOURCE_ALARM

YES

Status Bit - IR Source Alarm

 

 

 

Logic High = IR Source output too

 

 

 

low

 

 

 

Logic Low = IR Source output

 

 

 

normal

22

ST_WHEEL_ALARM

YES

Status Bit - Wheel Temperature Alarm

 

 

 

Logic High = Wheel Temp out of spec

 

 

 

Logic Low = Wheel Temp in spec

23

CO_MEASURE

NO

IR detector reading during measure

 

 

 

phase. Typically 2500-4500 mV

24

CO_REFERENCE

NO

IR detector reading during reference

 

 

 

phase. Typically 2500-4500 mV

25

SAMPLE_PRES

NO

Sample pressure in mV. Typical sea

 

 

 

level value = 4300mV for 29.9" HG-

 

 

 

A.

26

SAMPLE_FLOW

NO

Sample flow in mV.

27

SAMPLE_TEMP

NO

Sample temp in mV

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 56

Page 62
Image 62
Teledyne 300 instruction manual St Flow Alarm

FAQ

What are the status bits for the Teledyne Model 300 CO Alarm?

The status bits include flow alarm, temperature alarm, diagnostic mode, power status, pressure alarm, high range, system status, bench temperature alarm, IR source alarm, and wheel temperature alarm.

How to interpret the logic levels for each status bit?

The logic high and low correspond to specific conditions for each status bit, indicating whether the measurement is within or out of spec.

Where can I find the P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual?

You can find the manual on page 56 of the official Teledyne website or by contacting their customer support.