WARNING MESSAGE CLEAR COMMANDS

WARNING MESSAGE CLEAR COMMANDS

Command

 

Warning Message Cleared

"W WSYSRES<CRLF>"

 

SYSTEM RESET

"W WRAMINIT<CRLF>"

 

RAM INITIALIZED

"W WSOURCE<CRLF>"

 

SOURCE WARNING

"W WBHEAT<CRLF>"

 

BENCH HEAT SHUTDOWN

"W WWHLTEMP<CRLF>"

 

WHEEL TEMP WARNING

"W WSMPFLOW<CRLF>"

 

SAMPLE FLOW WARNING

"W WSMPRES<CRLF>"

 

SAMPLE PRESSURE WARNING

"W WSMPTEMP<CRLF>"

 

SAMPLE TEMP WARNING

"W WBOXTEMP<CRLF>"

 

BOX TEMP WARNING

"W WBNCHTEMP<CRLF>"

 

BENCH TEMP WARNING

"W WDYNZERO<CRLF>"

 

CANNOT DYN ZERO

"W WDYNSPAN<CRLF>"

 

CANNOT DYN SPAN

"W WVFINS<CRLF>"

 

V/F NOT INSTALLED

 

TABLE 7.2

7.3 Status/control

This subset of messages is concerned with reporting the status of the Analyzer and controlling the Analyzer remotely. Whenever the Analyzer does a calibration it issues a report to the RS-232 output. The table on the following page summarizes the status reports.

STATUS REPORTS

Report

"C DDD:HH:MM IIII START ZERO CALIBRATION" "C DDD:HH:MM IIII FINISH ZERO CALIBRATION" "C DDD:HH:MM IIII START SPAN CALIBRATION" "C DDD:HH:MM IIII FINISH SPAN CALIBRATION"

"C DDD:HH:MM IIII START MULTI-POINT CALIBRATION" "C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION"

"C DDD:HH:MM IIII START CALIBRATION HOLD" "C DDD:HH:MM IIII FINISH CALIBRATION HOLD"

TABLE 7.3

To do a remote adjustment via the RS-232 interface, the host computer should issue a message with the following format:

"C COMMAND<CRLF>

The commands are summarized in the table below:

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 64

Page 70
Image 70
Teledyne 300 instruction manual Status/control, Warning Message Clear Commands, Warning Message Cleared, Status Reports

FAQ

What are the warning message clear commands?

The warning message clear commands for the Teledyne API Model 300 CO Analyzer can be found in the official instruction manual on Page 64.

How can I perform remote adjustments via the RS-232 interface?

To perform remote adjustments via the RS-232 interface, the host computer should issue a message with a specific format, as detailed in the instruction manual.

Where can I find the status reports for the Analyzer?

The status reports for the Teledyne API Model 300 CO Analyzer can be found in the official instruction manual on the following page.