WARNING MESSAGE CLEAR COMMANDS

WARNING MESSAGE CLEAR COMMANDS

Command

 

Warning Message Cleared

"W WSYSRES<CRLF>"

 

SYSTEM RESET

"W WRAMINIT<CRLF>"

 

RAM INITIALIZED

"W WSOURCE<CRLF>"

 

SOURCE WARNING

"W WBHEAT<CRLF>"

 

BENCH HEAT SHUTDOWN

"W WWHLTEMP<CRLF>"

 

WHEEL TEMP WARNING

"W WSMPFLOW<CRLF>"

 

SAMPLE FLOW WARNING

"W WSMPRES<CRLF>"

 

SAMPLE PRESSURE WARNING

"W WSMPTEMP<CRLF>"

 

SAMPLE TEMP WARNING

"W WBOXTEMP<CRLF>"

 

BOX TEMP WARNING

"W WBNCHTEMP<CRLF>"

 

BENCH TEMP WARNING

"W WDYNZERO<CRLF>"

 

CANNOT DYN ZERO

"W WDYNSPAN<CRLF>"

 

CANNOT DYN SPAN

"W WVFINS<CRLF>"

 

V/F NOT INSTALLED

 

TABLE 7.2

7.3 Status/control

This subset of messages is concerned with reporting the status of the Analyzer and controlling the Analyzer remotely. Whenever the Analyzer does a calibration it issues a report to the RS-232 output. The table on the following page summarizes the status reports.

STATUS REPORTS

Report

"C DDD:HH:MM IIII START ZERO CALIBRATION" "C DDD:HH:MM IIII FINISH ZERO CALIBRATION" "C DDD:HH:MM IIII START SPAN CALIBRATION" "C DDD:HH:MM IIII FINISH SPAN CALIBRATION"

"C DDD:HH:MM IIII START MULTI-POINT CALIBRATION" "C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION"

"C DDD:HH:MM IIII START CALIBRATION HOLD" "C DDD:HH:MM IIII FINISH CALIBRATION HOLD"

TABLE 7.3

To do a remote adjustment via the RS-232 interface, the host computer should issue a message with the following format:

"C COMMAND<CRLF>

The commands are summarized in the table below:

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 64

Page 70
Image 70