BENCH HEAT

Temperature control of

Check Optical Bench

SHUTDOWN

the Optical bench cannot

heater and thermistor as

 

be maintained at its

described in Section 9.10

 

48oC set point

 

SAMPLE PRESSURE

The Sample Pressure is

Check for pressure

WARNING

less than 15"Hg or is

transducer problems as

 

greater than 35"Hg

described in Sect 10.6.5

 

 

and Sect 10.6.1

SAMPLE FLOW

The sample flow is less

Check for pneumatic

WARNING

than 500 cc/min or

system problems as

 

greater than 1000

described in Section

 

cc/min.

10.6.1 Check for flow

 

 

transducer problems as

 

 

described in Sect 10.6.5

BOX TEMP

The inside chassis temp

See Section 10.6.2

WARNING

is less than 10oC or is

 

 

greater than 50oC

 

SAMPLE TEMP

The Sample

See Section 10.6.2

WARNING

Temperature is less than

 

 

10oC or is greater than

 

 

50oC

 

CANNOT DYN ZERO

An offset of more than

See Section 10.5.5

 

+5 ppm would be

 

 

required to Zero adjust

 

 

the analyzer

 

CANNOT DYN SPAN

A slope of less than 0.5

See Section 10.5.6

 

or greater than 2.0 would

 

 

be required to Span

 

 

adjust the analyzer

 

V/F NOT

The CPU is unable to

Check and re-seat CPU

INSTALLED

communicate with the

and V/F board. See

 

V/F Board

Sect. 10.6.3

SYSTEM RESET

A power Off-On cycle

None required

 

has occurred

 

RAM INITIALIZED

Dynamic memory has

None required

 

been re-initialized in

 

 

response to the

 

 

installation of a new

 

 

PROM or memory chip

 

 

TABLE 10.1

 

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 86

Page 92
Image 92
Teledyne 300 instruction manual Bench Heat

FAQ

How do I troubleshoot sample pressure issues?

Check for pressure transducer problems as described in Section 10.6.5 and Section 10.6.1.

What should I do if the sample flow is too low or too high?

Check for pneumatic system problems as described in Section 10.6.1 and Section 10.6.5. Check for flow transducer problems as described in Section 10.6.5.

How can I fix communication issues with the V/F board?

Check and re-seat the CPU and V/F board. See Section 10.6.3 for more information.