Teledyne 300 instruction manual Push “CAL” 15.Enter password. If enabled

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 81
Image 81

12.Set the calibrator to produce 75% to 85% of the URL (upper range limit). This will be 40 ppm ±2.5 ppm on the 50 ppm range.

13.Allow the Analyzer to sample the CO concentration.

14.Push “CAL”

15.Enter password. (If enabled)

16.Push “CONC”

17.Change the span value in the display to the calculated CO concentration in the manifold in ppm units, and push “ENTR”.

18.Wait 15 minutes for the Analyzer to stabilize.

19.Push “SPAN” and “ENTER.” The Analyzer is now spanned. Record the calculated CO concentration, the DVM reading and the percentage chart reading, or the RS-232 output.

20.Push “EXIT,” “EXIT.” The Analyzer is now returned to the normal Sample mode.

21.Introduce at least five (5) more approximately evenly spaced CO concentrations into the manifold to complete the manual calibration.

22.Record all calculated CO concentrations, DVM reading and strip-chart recorder readings or RS-232 output readings.

23.Plot the calculated CO concentrations (X-axis) versus output voltages and/or percentage chart readings (Y-axis).

Calculate the curve equations:

[CO] ppm = (Volts - b) / m

where b is the offset (should be within ± 0.05 volts of zero setting)and m is the slope (should be .098 to .102 based on 0-5V full scale)

or

[CO] = (% chart - b) / m

(b should be within ± 1% of chart and m should be 1.96 to 2.04)

The correlation coefficient r should be 0.998 or higher.

The Analyzer is now calibrated. All CO concentration data should be obtained by reading the Analyzer output in volts or percentage of chart and converting to ppm from the appropriate equation or curve.

If poor correlation exists, check for

1.Flow calculation errors,

2.Concentration calculation errors,

3.Leaks in manifold,

4.Dirt in the manifold,

5.Proper manifold venting,

6.Zero air system.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 75

Page 81
Image 81
Teledyne 300 instruction manual Push “CAL” 15.Enter password. If enabled

FAQ

What is the first step in calibrating the Teledyne API Model 300 CO Analyzer?

The first step is to set the calibrator to produce 75% to 85% of the upper range limit.

How long should I wait for the Analyzer to stabilize after changing the span value?

Wait 15 minutes for the Analyzer to stabilize before proceeding.

What should I do if poor correlation exists after calibration?

Check for flow calculation errors, concentration calculation errors, leaks in the manifold, dirt in the manifold, proper manifold venting, and zero air system.