Teledyne 300 5.2.3 D/A output, Electric Test, 2 I/O SIGNALS, 0%, 20%, 40%, 60%, 80%, 100%, 0%

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 63
Image 63

28

BENCH_TEMP

NO

Optical Bench temp. Typically 2270

 

 

 

mV for 48 C

29

WHEEL_TEMP

NO

Filter Wheel temp. Typically 4770

 

 

 

mV for 68 C

30

BOX_TEMP

NO

Internal analyzer temp in mV

31

DCPS_VOLTAGE

NO

DC power supply composite voltage

 

 

 

output. Typically 2500 mV.

32

DAC_CHAN_0

NO

Output of DAC 0(REC) in mV.

33

DAC_CHAN_1

NO

Output of DAC 1(DAS) in mV.

34

DAC_CHAN_2

NO

Output of DAC 2(TEST) in mV.

35

DAC_CHAN_3

NO

Output of DAC 3(Spare) in mV.

36

CO_CONC_1

YES

CO Reading (REC) in mV

37

CO_CONC_2

YES

CO Reading (DAS) in mV

38

TEST_OUTPUT

YES

Test Channel in mV

TABLE 5.2 I/O SIGNALS

5.2.3 D/A output

This test cycles 3 of the analog output channels from 0% to 100% of Full Scale in 20% FS steps. It starts by outputting 0 volts to all four channels and displaying a 0% button. Then, every five seconds, the output is increased 20% FS and the button is changed accordingly. Thus, the button (and the analog outputs) will cycle through the following value

0%, 20%, 40%, 60%, 80%, 100%, 0%, ...

To pause the output at the current voltage, press the n% button. To resume automatic cycling, press the n% button again.

5.2.4 Electric Test

This test activates a diagnostic circuit located on the Synchronous Demodulator board which generates an artificial signal which simulates the output of the IR detector. This signal is injected in place of the detector output.

When activated, Electric Test will automatically switch the analyzer into a 50 ppm range and result in the Analyzer producing a constant, stable output (i.e. CO reading) of about 40 ppm. This test is particularly useful in isolating problems, since it exercises essentially all electronic sub-systems of the Analyzer but does not depend on the proper function of Optical or Pneumatic subsystems.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 57

Page 63
Image 63
Teledyne 300 instruction manual 5.2.3 D/A output, Electric Test, 2 I/O SIGNALS, 0%, 20%, 40%, 60%, 80%, 100%, 0%

FAQ

What is the purpose of the Electric Test?

The Electric Test activates a diagnostic circuit to simulate the output of the IR detector, providing a constant, stable CO reading of about 40 ppm.

How does the D/A output test work?

The D/A output test cycles 3 analog output channels from 0% to 100% of Full Scale in 20% FS steps, exercising electronic sub-systems of the Analyzer.

What are the I/O signals for the Teledyne API Model 300 CO Analyzer?

The I/O signals include bench, wheel, and box temperatures, DC power supply composite voltage, and various output readings in mV.