Teledyne 300 instruction manual Adjustments, Power supply board adjustment, Box temperature limits

Models: 300

1 121
Download 121 pages 3.98 Kb
Page 83
Image 83
9.0 ADJUSTMENTS

9.0 ADJUSTMENTS

NOTE

THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!

All adjustments to the Model 300 are easy to make. Pots and test points are readily accessible without removing any components.

Figure 1.6 is a plan view of the Model 300 CO Analyzer showing all the major components.

Figure 9.1 is an electrical diagram of Model 300 CO Analyzer.

9.1 Power supply board adjustment

The power supply board provides +/-15v +12v and +5v DC power to the Analyzer. Four temperature linearization circuits, for the Analyzer main lamp, Sample temp, Box temp, and IZS carbon monoxide generator are also located on the power supply board.

Each circuit is a whetstone bridge with the measuring thermistor being one leg. A feedback circuit performs the required linearization. Zero adjust pots have been factory set and no field adjustment should be required.

9.1.1 Box temperature limits

The box temperature is measured by a thermistor located on the motherboard. The box temp is not controlled in the Model 300. The temperature is measured and displayed as a TEST function on the front panel (see Section 5.1). The alarm limits can be set via an RS-232 port command.

P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 77

Page 83
Image 83
Teledyne 300 instruction manual Adjustments, Power supply board adjustment, Box temperature limits

FAQ

What adjustments can be made to the Model 300?

Qualified maintenance personnel can make adjustments to the power supply board and box temperature limits.

Where are the adjustment points located?

The adjustment points are easily accessible without removing any components.

Are field adjustments required for the zero adjust pots?

The zero adjust pots have been factory set and should not require any field adjustment.